Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual
Page 9

IM 12D08B02-01E
CONFIGURATION CHECKLIST FOR SC202
Primary choices
default
alternatives
reference on page
menu
Measurement
Conductivity
Resistivity
5.8- 5.9
SC 01
Range
0-1000 μS/cm
max. 1999 mS/cm
5.3
“range”
Temperature unit
Celsius
Fahrenheit
5.10- 5.11
SC 11
Sensor
Cell constant
0.1 /cm
any value between 0.08
and 50
5.8-5.9, 6.1- 6.3
SC 03
Sensor type
2-electrode
4- electrode
5.8- 5.9
SC 02
Temperature compensator
Pt1000
Ni100, Pt100, 8k55, Pb36
5.10-5.11
SC 10
Choices
Communication
enabled
disable HART
(R)
, PH201*B
5.19
SC 60- 62
Burn out
inactive
HI or LO output on fail
5.14- 5.15
SC 32
Temperature compensation
NaCl in water
fixed T.C., matrix
5.12, 5.13, 5.5
SC 20- 28; “temp”
USP functionality
inactive
Fail if USP limits are
9.1, 9.2, 5.17
SC 57
exceeded
HOLD during maintenance
inactive
HOLD last value or fixed
value
5.17, 5.3- 5.4
“hold”, SC 50
Calibration temperature
inactive
adjustment +/- 15 °C
5.11
SC 12
ZERO calibration
inactive
adjustment +/-1 μS/cm
5.9
SC 04
Diagnostics
hard alarm on hard or soft choices
5.17
SC 53
all errors
Cell fouling alarm
active
except E13 inactive
5.9
SC 05
Password protection
inactive
password for different levels
5.17
SC 52
Output in Concentration units
inactive
linearization of output, w%
5.14 - 5.17
SC 31/35/55
on LCD