Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual
Page 63

IM 12D08B02-01E
Parameter setting 5-17
Code Display
Function
Function
detail
X
Y
Z
Default
values
User interface
50
*RET.
Auto return
Auto return to measuring mode Off
0
Auto return to measuring mode On
1
1
On
51
Not
used
52 *PASS Passcode
Maintenance
passcode
Off
0 0.0.0
Off
Note # = 0 - 9, where Maintenance passcode On
#
Commissioning passcode Off
0
Off
1=111, 2=333, 3=777 Commissioning passcode On
#
4=888, 5=123, 6=957 Service passcode Off
0
Off
7=331, 8=546, 9=847 Service passcode On
#
53
*Err.01
Error setting
Polarization too high
Soft/Hard
0/1
1
Hard
*Err.05
Shorted
measurement
Soft/Hard
0/1
1
Hard
*Err.06
Open
measurement Soft/Hard
0/1
1
Hard
*Err.07
Temperature sensor open Soft/Hard
0/1
1
Hard
*Err.08
Temp. sensor shorted
Soft/Hard
0/1
1
Hard
*Err.13
USP limit exceeded
Soft/Hard
0/1
0
Soft
54
*E5.LIM
E5 limit setting
Maximum conductivity value
250 mS
(Minimum resistivity value)
0.004 kΩ
*E6.LIM
E6 limit setting
Minimum conductivity value
1.000 μS
(Maximum resistivity value)
1.000 MΩ
55
*%
Display mA in w/w% mA-range displayed in w/w% off
0
0
Off
mA-range displayed in w/w% on
1
*0%
Set 0% output value in w/w%
*100%
Set 100% output value in w/w%
56
*DISP.
Display resolution
Auto ranging display
0
0
Auto
Display fixed to X.XXX μS/cm or MΩ·cm 1
Display fixed to XX.XX μS/cm or MΩ·cm 2
(2)
Display fixed to XXX.X μS/cm or MΩ·cm 3
Display fixed to X.XXX mS/cm or kΩ·cm 4
Display fixed to XX.XX mS/cm or kΩ·cm 5
Display fixed to XXX.X mS/cm or kΩ·cm 6
Display fixed to XXXX mS/cm or kΩ·cm 7
57
*USP.
USP setting
Disable the E13 (USP limit exceeded)
0
0
Off
Enable the E13 (USP limit exceeded)
1
58-59
Not
used