Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual
Page 67

IM12B07D02-01E
5-16 Parameter setting
Code Display
Function
Function detail
X Y
Z Default values
Calibration functions
20 *∆t.SEC
Stability check time
5 sec.
*∆PH
Stability check pH
0.02 pH
21
*AS.LOW
As Pot low limit
-120 mV
(As Pot) *AS.HI
As Pot high limit
120 mV
21
*ZP.LOW
Zero Point low limit
5.00 pH
(Zero) *ZP.HI
Zero Point high limit
9.00 pH
22
*SL.LOW
Slope low limit
70 %
*SL.HI
Slope high limit
110 %
23
*ITP
Set ITP
Preset calibration data from manufacturer
7.00 pH
(pH)
*SLOPE
Set slope
or from laboratory determinations.
100 %
*ASP.1D
Set As Pot
For the main parameter
0.0 mV
Press YES to confirm 0.1 mV resolution,
then set value with >, ^, ENT keys.
Press NO to change to *ASP.
*ASP
Set As Pot
For the main parameter
Press YES to confirm 1 mV resolution,
then set value with >, ^, ENT keys.
*ASPmV
Set As Pot ORP
For parameter 2 (when activated in
service code 02)
23
*ASP.1D
Set As Pot (ORP)
For the main parameter
(ORP)
Press YES to confirm 0.1 mV resolution,
then set value with >, ^, ENT keys.
Press NO to change to *ASP.
*ASP
Set As Pot
For the main parameter
Press YES to confirm 1 mV resolution,
then set value with >, ^, ENT keys.
24
*BUF.ID
Buffer table 4
Buffer tables to NIST (formerly NBS)
25
*BUF.ID
Buffer table 7
(see section 10 for table details)
26
*BUF.ID
Buffer table 9
User adjustable for special requirements
27
*ZERO.P
Enable zero point in Disable zero point (enable As Pot)
0
0 Disabled
pH units
Enable zero point (disable As Pot)
1
28-29
Not
used
Fig. 5-1.
pH
mV
ITP
As pot
0 mV
0 °C
10 °C
25 °C
mV
500
0
- 200
14
pH
ITP
Zero Point
0
7
Select YES/NO
Select YES/NO