Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual
Page 65

IM12B07D02-01E
5-14 Parameter setting
Code Display
Function
Function
detail
X
Y
Z
Default
values
Temperature measuring and compensation functions
10 *T.SENS Temperature
sensor
Pt1000
0 0 Pt1000
Pt100
1
3kBalco
2
5k1
3
8k55
4
350
5
6k8
6
PTC10k
7
11
*T.UNIT
Display in °C or °F
°C
0
0
°C
°F
1
12
*T.ADJ
Calibrate temperature Adjust to allow for cable resistance
None
13
*T.COMP
Set temp comp
Compensation for process changes off
0
0
Off
Compensation for process changes on
1
*T.COEF
Adjust process TC
Set for TC in pH per 10 °C
-0.00 pH per
10 °C
14-19
Not
used
See also other documents in the category Yokogawa Sensors:
- EJA438 (5 pages)
- EJA120A (6 pages)
- EJA115 (85 pages)
- EJA120A (47 pages)
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA210A (70 pages)
- EJA130A (4 pages)
- EJA430A (78 pages)
- EJA210E (96 pages)
- EJA210E (52 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJX910A (175 pages)
- EJX910A (83 pages)
- EJX910A (9 pages)
- EJX910A (103 pages)
- FlowNavigator Software (163 pages)
- EJX910A (55 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (85 pages)
- EJA110E (4 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJX430A (40 pages)
- EJX430A (76 pages)
- EJA430E (7 pages)
- EJX430A (6 pages)
- EJA430E (41 pages)
- EJA430E (96 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)