IM 12B07D02-01E

Operation 4-1

4. OPERATION; DISPLAY FUNCTIONS AND SETTING

4-1. Operator interface

This section provides an overview of the operation of the EXA operator interface. The basic procedures

for obtaining access to the three levels of operation are described briefly. For a step-by-step guide to

data entry, refer to the relevant section of this user’s manual. Figure 4-1 shows the EXA operator inter-

face.

LEVEL 1: Maintenance

These functions are accessible by pushbutton through a flexible front cover window. The functions make

up the normal day-to-day operations that an operator may be required to complete. Adjustment of the

display and routine calibration are among the features accessible in this way. (See table 4-1).

LEVEL 2: Commissioning

A second menu is exposed when the EXA front cover is removed and the display board is revealed.

Users gain access to this menu by pressing the button marked

*

in the lower right of the display board.

This menu is used to set such values as the output ranges and hold features. It also gives access to the

service menu. (See table 4-1).

LEVEL 3: Service

For more advanced configuration selections, press the button marked

*

, then press “NO” repeatedly

until you reach *SERV. Now push the “YES” button. Selecting and entering “Service Code” numbers in

the commissioning menu provide access to the more advanced functions. An explanation of the Service

Codes is listed in chapter 5 and an overview table is shown in chapter 10.

Table 4-1. Operations overview

Routine

Function

Chapter

Maintenance

AUT.CAL

Calibration with programmed buffer solutions

6

MAN.CAL

Calibration with other buffer solutions

6

SAMPLE

Grab

sample

calibration

6

DISP

Read auxiliary data or set message display

4

MAN.IMP

Manual

start

of

impedance

check

5

TEMP

Select automatic or manual compensation

5

HOLD

Switch

hold

on/off

(when

activated)

5

Commissioning

*OUTP

Adjust

the

output

range

5

*HOLD

Activate

the

hold

function

5

Service

*SERV

Fine tune the specialized functions of the transmitter

5

(Access to coded entries
from the commissioning
level)

NOTE:

All three levels may be separately protected by a password. See Service Code 52 in chapter 5

Service Code table for details on setting passwords.