5.2 DEFINITIONS
GENERAL SETUP
Display Contrast
It is possible to adjust the contrast of the
display using the key to give optimum
viewability.
Total Units
Enables totals to be integrated at 1/1000 of the
flowrate by programming x 1000. For
example, if the flowrate is in scm/h the total
will be kscm/h (scm x 1000).
Timebase
Timebase selection will determine whether the
rate of flow is displayed in kg/day,
kg/hour...etc.
Front Access
If enabled, access to the Calibration routine is
possible via the front panel (via the Total and
Display keys). If disabled, access to the
Calibration routine is only possible by
connecting a link between terminals 1 and 2.
Front Reset
If disabled, the front panel reset key becomes
inoperable during normal operation.
GAS PARAMETERS
Default
Display
The flowrate and total which is normally
displayed (eg. Mass or Corrected Volume).
Calibration 59