Technical data, Technical data -1, Ii 2 g – Yokogawa RAGN User Manual
Page 33: Option specifications, Standard specifications

<8. TECHNICAL DATA>
8-1
IM 01R01B10-00E-E 2nd edition: November 01, 2011-00
All Rights Reserved. Copyright © 2011, Rota Yokogawa
8. Technical Data
OPTION SPECIFICATIONS
Limit switch (option /GM1 to /GM5):
( for P- tubes with PVDF- or SS- float with magnet only)
Type
: reed contact with bistable switching
Max. switching voltage : 230 V
Max. switching current : 2 A
Max. switching capacity : 40 W/VA
Temperature range : -10°C to +70°C
Protection
: IP65
Internal capacity
: 0 nF
Internal inductivity : 0 mH
Electrical connection : LIYY 2 x 0.34 mm²; length 1 m
Housing
: Polystyrene
Weight
: 35 g
Explosion proof
:
Intrinsic safe acc. EN 60079-11 chapter 5.7,
IEC 60079-11 chapter 5.7 and ANSI/ISA 60079-11
chapter 5.7 as "Simple Apparatus".
Group
: IIC
Category
: 2G
Temperature class : T6
Entity parameter : Ui = 15 V ; Ii = 50 mA ; Pi = 187 mW
Li ≈ 0 mH ; Ci ≈ 0 nF
Limit switch (option /GR2 to /GR8):
(for L- tubes with PVDF float only)
Type
: Bistable inductive ring sensor
Power supply
: 4.5 V to 15 V DC
Consumption
: acc. DIN EN 60947-5-6 (NAMUR)
Float below ring sensor : < 1 mA
Float above ring sensor : > 2.2 mA
Temperature range : -25°C to +65°C non-Ex- type
Protection
: IP 67
Electrical connection : 2 x 0.14 mm² , with shield 0.4 mm²,
2 m long
Explosion proof type (option /KS1):
Temperature range : -25°C to +60°C
Marking acc. guideline 94/9/EG :
Manufacturer : Rota Yokogawa, Rheinstr.8,
D-79664 Wehr
Type
: RI20-10K/G or RI20-17K/G
Year of production : in serial number
Protection
: Ex ia
Group
:
IIC
Category
: 2G
Temperature class : T6
Certificate No. : PTB 03ATEX 2111
Safety relevant data (see also certificate for data):
Ui = 12 V, Ii = 22 mA , Pi = 66 mW,
Li = 20 mH, Ci = 200 nF
CE-marking
:
II 2 G
Power supply for limit switch (option /W__):
Type : acc. DIN EN 60947-5-6 (NAMUR)
- KFA5-SR2-Ex*-W (115 V AC), * = 1 or 2
- KFA6-SR2-Ex*-W (230 V AC), * = 1 or 2
- KFD2-SR2-Ex*-W (24 V DC), * = 1 or 2
- KHA6-SH-Ex1 (115/230 V AC), Fail Safe, 1 channel
- KFD2-SH-Ex1 (24 V DC), Fail Safe, 1 channel
Power supply :
- 230 V AC ± 10%, 45-65Hz
- 115 V AC ± 10%, 45-65Hz
- 24 V DC ± 25%
Relay output :
1 or 2 potential-free changeover contact(s)
Switching capacity :
max. 250 V AC, max. 2 A
STANDARD SPECIFICATIONS
Fluids to be measured : Liquids and gas
Measurable flow rates
:
- Water (20 °C)
: 0.002 l/h to 10 m3/h
- Air (20 °C; 1 bar abs.) : 0.1 l/h to 160 m3/h
Turndown
:
- P metering tube
: 10:1
- L metering tube
: 20:1
Metering tubes
: L6; L7; P0; P1; P2; P4 (length 300 mm)
Process temperature
: -25°C to +100°C
Process pressure
:
Metering tube
L6;L7
P0;P1
P2
P4
Pmax (bar)
16
10
8
6
Installation length
:
Process connection
Thread
Clamp
Flange
Length [mm]
375
375
425
Weight
: Depending on design (see page 7)
Accuracy
:
Tube
Measuring accuracy acc.
Directive VDI/VDE 3513 sheet 2 (q
G
= 50%)
L613 - L623
2.5%
L624 - L747
1.6%
P051 - P471
1.6%
Materials
:
Threads G, NPT
: AISI 316L (1.4404)
Flange EN / ASME
: AISI 316L (1.4404)
Clamp ISO 2852
: AISI 316L (1.4404)
Housing
: AISI 304 (1.4301)
Nut
: AISI 316 (1.4401) (or galvanized steel)
Stoppers (L6, L7 tube) : PFA
Stoppers (P0 - P4 tube) : PVDF, AISI 316L (1.4404)
Measuring cone
: Borosilicate glass
Float (L6, L7 tube )
: Titanium, PVDF
Float (P0 - P4 tube)
: PTFE, PVDF (FDA conform),
AISI 316Ti (1.4571)
Gaskets
: NBR, FKM,EPDM (FDA conform)
Pressure Equipment Directive (PED) Directive 97/23/EG:
Models
: RAGN04, RAGN05, RAGN06
Tubes :
- Modul
: A
- Fluid Group
: 1 (liquid, gases)
- Produced acc. to category
: I
FDA-Conformity:
RAGN with P- tube, PVDF- or SS- float and EPDM- gaskets
(option /ME).
Stoppers and floats made of PVDF:
21 CFR § 177 2510(a)
O-rings made of EPDM:
21 CFR § 177 2600-21
Compliance with safety application acc. IEC 61508: 2010
and ISO 13849:
Please refer to the FMEDA report and instruction manual.