4 easy setup, 1 tag number (tag), 2 calibration range (scale in) and pressure unit – Yokogawa EJA115 User Manual
Page 28: Easy setup -6, Tag number (tag) -6, Calibration range (scale in) and pressure unit -6
IM 01C22T03-00E
5-6
5. CONFIGURATION
Enter [Old Address] to 5 and [New Address] to 20
and Click [Set]
“New address successfully set” is displayed in the State
field.
Click [Close]
Figure 5.9 Change the address
When changing a parameter after changing the Ad-
dress, save the data and close the DTM Works win-
dow.
Select Open Device Maintenance Info from the Action
pull-down menu on the toolbar.
Change the address of Basic Information and return to
the DTM Works window.
5.4 Easy Setup
Easy Setup tracks parameters that often need to be
configured or changed.
In the DTM Works window, click [Device]
→ [Con-
nect]
Click [Device]
→ [Parameter] → [Online parameter]
Click on [Easy Setup] folder
IMPORTANT
Do not turn power OFF immediately after param-
eter setting. When parameters are saved to the
EEPROM, the redundant processing is executed
for an improvement of reliability. Should the
power be turned OFF within 60 seconds after
setting of parameters, changed parameters are
not saved and may return to their original values.
Note.1 The following table shows which operation
modes are supported by the blocks.
Table 5.4 Operation Mode
T0504.EPS
AI Block Transducer Block Physical Block
Automatic (AUTO)
Yes
Yes
Yes
Manual (MAN)
Yes
N/A
N/A
Out of Service (O/S) Yes
Yes
Yes
Refer to the “APPENDIX 1. LIST OF PARAM-
ETERS FOR EACH BLOCK OF THE EJA” for details
of the Write Mode for each block.
5.4.1 Tag Number (TAG)
Enter the Tag number in TAG.
Press the Enter key to confirm the data.
Figure 5.10 Tag Number Setup
5.4.2 Calibration Range (Scale In) and
Pressure Unit
Change the Target Mode of Transducer Block to Out
of Service (O/S).
AL.22 occurs when Target Mode is changed to Out of
Service (O/S).
Figure 5.11 Target Mode Setup