Blood pressure measurements determined with this device are equivalent to those obtained by
a trained observer using the cuff/stethoscope auscultation method, within the limits prescribed
by the American National Standard, Electronic or Automated Sphygmomanometers (AAMI
SP10). To receive a copy of the report containing the AAMI SP10 test results, contact ZOLL
Technical Support at (800) 348-9011 or (978) 421-9655
.
Operating Time (with SpO
2
, EtCO
2
and NIBP options)
For a new, fully charged lithium ion battery pack at 20°C (68°F):
•
90 defibrillator discharges at maximum energy (200 J), or
•
2.75 hours minimum of continuous ECG monitoring and blood
pressure measurements once every 5 minutes, or 2.25 hours of
continuous ECG monitoring/pacing at 60 mA, 70 beats per
minute.
Environmental
Operating Temperature: 0
to 40 C (32º to 104º F)
Storage Temperature: -20
to 60 C (-4º to 140º F)
Humidity: 5 to 95% relative humidity, non-condensing
Electromagnetic Immunity
AAMI DF-80; EN 61000-4-3: 2002, 10 V/m
Operating Pressure
594 to 1060 mBar