Specifications – ZOLL R Series Monitor Defibrillator Rev A NIBP User Manual
Page 27
Specifications
9650-0914-01 Rev. A
R Series Non-Invasive Blood Pressure (NIBP)
25
Specifications
Principle of Operation
Oscillometric
Warm-up Time
Operational in less than 10 seconds
Regulatory Standards
ANSI/AAMI SP10-2002;
EN 1060-1: 1996 Specification for Non-invasive sphygmomanometers;
EN 1060-3: 1997;
EN 60601-2-30: 2000
NIBP Measurement Range
Systolic
Adult: 40 – 260 mmHg; 5.3 – 34.6 kPa
Pediatric: 40 – 160 mmHg; 5.3 – 21.3 kPa
Neonatal: 40 – 130 mmHg; 5.3 – 17.3 kPa
Diastolic
Adult: 20 – 200 mmHg; 2.7 – 26.7 kPa
Pediatric: 20 – 120 mmHg; 2.7 – 16.0 kPa
Neonatal: 20 – 100 mmHg; 2.7 – 13.3 kPa
Mean
Adult: 30 – 220 mmHg; 4.0 – 29.3 kPa
Pediatric: 30– 133 mmHg; 4.0 – 17.7 kPa
Neonatal: 30 – 110 mmHg; 4.0 – 14.6 kPa
NIBP Measurement Accuracy
Per AAMI SP10 +/- 5 mmHg (0.7 kPa) mean difference;
8 mmHg (1.1 kPa) standard deviation
a
NIBP Measurement Resolution
1 mmHg; 0.13 kPa
NIBP Measurement Cycle Time
Typical: 30
seconds
Worst Case:
180 seconds
Pulse Rate Range
30 – 220 pulses per minute
Pulse Rate Accuracy
±2% or ±3 bpm, whichever is greater
Pulse Rate Resolution
1 bpm (beats per minute)
Alarm Limits
Systolic
Adult
High 80 – 260 mmHg;
High 10.7 – 34.7 kPa
Low 40 – 140 mmHg;
Low 5.3 – 18.7 kPa
Pediatric
High 80 – 160 mmHg;
High 10.7 – 21.3 kPa
Low 35 – 140 mmHg;
Low 4.7 – 18.7 kPa
Neonatal
High 60 – 130 mmHg;
High 8.0 – 17.3 kPa
Low 25 – 120 mmHg;
Low 3.3 – 16.0 kPa
Diastolic
Adult
High 50 – 200 mmHg;
High 6.7 – 26.7 kPa
Low 25 – 100 mmHg;
Low 3.3 – 13.3 kPa
Pediatric
High 50 – 130 mmHg;
High 6.7 – 17.3 kPa
Low 20 – 100 mmHg;
Low 2.7 – 13.3 kPa
Neonatal
High 30 – 105 mmHg;
High 4.0 – 14.0 kPa
Low 20 – 100 mmHg;
Low 2.7 – 13.3 kPa
Mean
Adult
High 60 – 220 mmHg;
High 8.0 – 29.3 kPa
Low 30 – 120 mmHg;
Low 4.0 – 16.0 kPa
Pediatric
High 60 – 140 mmHg;
High 8.0 – 18.7 kPa
Low 30 – 120 mmHg;
Low 4.0 – 16.0 kPa
Neonatal
High 35 –110 mmHg;
High 4.7 – 14.7 kPa
Low 30 – 105 mmHg;
Low 4.0 – 14.0 kPa