Setting Alarms
The R Series NIBP option includes blood pressure alarms for the measurements shown in the
following table.
Before taking a measurement, ensure that all the NIBP alarms settings are appropriate for the
situation. See the R Series Operator’s Guide for instructions on setting alarms. When an NIBP
or heart rate alarm is triggered, the unit automatically initiates a single blood pressure
measurement. See the R Series Configuration Guide for instructions on how to enable/disable
this automatic measurement.
When the R Series unit is turned on, all alarm functions are disabled, and the alarm limits are
set to their default values for the selected patient type. Any changes to these settings remain in
effect until either the settings are changed for the patient type or for 10 seconds after the
R Series unit is turned off. You can configure the default alarm limits at power-up; see the
R Series Configuration Guide for more information.
Setting Alarm Limits
You can change all the alarm limits using standard R Series procedures. See the R Series
Operator’s Guide for instructions on setting alarm limits.
Patient Type
Alarm Parameter
Default Setting
Range (increments of 5)
Adult
Systolic
High
160 mmHg (21.3 kPa)
80-260 mmHg (10.7-34.7 kPa)
Low
90 mmHg (12.0 kPa)
40-140 mmHg (5.3-18.7 kPa)
Diastolic
High
110 mmHg (14.7 kPa)
50-200 mmHg (6.7-26.7 kPa)
Low
50 mmHg (6.7 kPa)
25-100 mmHg (3.3-13.3 kPa)
Mean
High
130 mmHg (17.3 kPa)
60-220 mmHg (8.0-29.3 kPa)
Low
60 mmHg (8.0 kPa)
30-120 mmHg (4.0-16.0 kPa)
Pediatric
Systolic
High
145 mmHg (19.3 kPa)
80-160 mmHg (10.7-21.3 kPa)
Low
75 mmHg (10.0 kPa)
35-140 mmHg (4.7-18.7 kPa)
Diastolic
High
100 mmHg (13.3 kPa)
50-130 mmHg (6.7-17.3 kPa)
Low
35 mmHg (4.7 kPa)
20-100 mmHg (2.7-13.3 kPa)
Mean
High
110 mmHg (14.7 kPa)
60-140 mmHg (8.0-18.7 kPa)
Low
50 mmHg (6.7 kPa)
30-120 mmHg (4.0-16.0 kPa)
Neonate
Systolic
High
100 mmHg (13.3 kPa)
60-130 mmHg (8.0-17.3 kPa)
Low
50 mmHg (6.7 kPa)
25-120 mmHg (3.3-16.0 kPa)
Diastolic
High
70 mmHg (9.3 kPa)
30-105 mmHg (4.0-14.0 kPa)
Low
30 mmHg (4.0 kPa)
20-100 mmHg (2.7-13.3 kPa)
Mean
High
80 mmHg (10.7 kPa)
35-110 mmHg (4.7-14.7 kPa)
Low
35 mmHg (4.7 kPa)
30-105 mmHg (4.0-14.0 kPa)