AED Pro Error Log Messages

9650-0309-01 Rev. A

ZOLL AED Pro Service Manual

2–7

14

Patient impedance experienced a short

15

The A2D is getting non-accurate readings during its test

17

The message queue overflowed

18

PowerOn task has timed out while waiting for a self test result to come back.

19

Calibration faults that occur during a clinical event.

20

An unexpected error has occurred when calling a Nucleus operation.

21

An unexpected error has occurred when calling a Rhapsody operation.

22

An unexpected coding error has occurred.

23

No Language installed - corrupt language.

24

The shock button is pressed/stuck.

25

The ECG Calibration data is invalid

26

The Patient Impedance Calibration data is invalid

27

A critical flash error has occurred

28

The WatchDog is not enabled during clinical event

29

The clock test failed

Note: Error numbers 256 and greater indicate noncritical conditions.

Table 2-4. AED Pro Error Log Messages (continued)

Error

No.

Error Message