0 manual self-test, 0 manual self-test -10 – ZOLL AED Pro Rev A User Manual
Page 22

C
HAPTER
1 M
AINTENANCE
T
ESTS
1–10
www.zoll.com
9650-0309-01 Rev. A
5.0 Manual Self-Test
Use this test to initiate the unit’s self-diagnostic procedure, which verifies the following:
•
Battery usage since last reset.
•
Defibrillation electrodes are properly connected.
•
ECG signal acquisition and processing electronics are functional.
•
Defibrillator charge and discharge circuitry (operational at 2 joules) is functional.
•
Microprocessor hardware and software are functional.
•
CPR monitoring circuitry and compression depth sensor are functional (when ZOLL
CPR-D•padz are attached).
•
Audio output circuitry is functional.
•
Display screen is functional.
Tools Needed
•
None.
Test Setup
•
Turn off the AED Pro unit.
•
If they are not already in place, preconnect defibrillation electrodes to the unit.
Test Procedure
Test
Action
Expected Result(s)
Pass
Fail
5.1
Press and hold the On/Off button
for at least 5 seconds.
• Within 10 seconds after the unit is
turned on, the Ready indicator
displays a green check.
;
UNIT OK
If an IrDA connection is established, turn off the unit; otherwise, the unit shuts down
automatically.