beautypg.com

0 battery check, 0 ecg monitoring (optional) – ZOLL AED Pro Rev A User Manual

Page 21

background image

Battery Check

9650-0309-01 Rev. A

ZOLL AED Pro Service Manual

1–9

3.0 Battery Check

Use this procedure to ensure that the battery shows no signs of damage or excessive wear.

Tools Needed

None.

Test Setup

None.

Test Procedure

4.0 ECG Monitoring (Optional)

Use this procedure to ensure that the ECG cable shows no signs of damage or excessive wear.

Tools Needed

Fluke Impulse 4000 Defibrillator/Pacer Performance Analyzer or equivalent

3-lead ECG cable

Test Setup

None.

Test Procedure

Test

Action

Expected Result(s)

Pass

Fail

N/A

3.1

Open battery door and verify that the battery has no signs of visible
damage.

‰

‰

‰

3.2

For a sealed lead acid battery, verify that the battery is not older
than 18 months (if it is older than 18 months, ZOLL recommends
that you replace the battery).

‰

‰

‰

Test

Action

Expected Result(s)

Pass

Fail

N/A

4.1

Insert the ECG cable into the connector on the AED Pro. Connect
the leads to the analyzer and set to NSR (normal sinus rhythm).

‰

‰

‰

4.2

Turn on the AED Pro unit.

Verify that the rhythm is
displayed.

‰

‰

‰

4.3

Grasp the ECG cable and leads
and gently move them back and
forth.

Verify that the signal is not
distorted.

‰

‰

‰