2.0 Preventive Maintenance Test

Use this test to verify that the AED Pro unit is working properly and ready for use.

For an alternative test using the AED Pro simulator, refer to the AED Pro Operator’s Guide.

Note: To verify that the delivered energy for the first, second, and third shocks corresponds

to the unit’s preconfigured adult energy levels, use the ZOLL Administration Software
(ZAS) to obtain the current configuration of the unit.

Tools Needed

Fluke Impulse 4000 Defibrillator/Pacer Performance Analyzer or equivalent

Defibrillation analyzer (universal) adapter cable

Test Setup

Turn off the AED Pro unit and unplug the preconnected electrodes. Keep the electrodes so
they can be reconnected later.

Connect the universal adapter cable to the AED Pro unit and to the analyzer.

Turn on the analyzer and set its function to VF (ventricular fibrillation).

Variations in Messages

When verifying messages in the test results, please be aware of the following issues:

The wording of some messages is configurable. The following table shows equivalent
messages.

The message that reports the detected electrode type can be either ADULT PADS,
PEDIATRIC PADS, or ECG CABLE.

Some messages are issued only if the unit is configured to do so. If an optional message is
not issued during a test, do not consider this a failure.

In some cases a text message on the display screen is abbreviated or otherwise differs
slightly from the associated voice message.

Default Message

Alternative Message

CHECK PATIENT

CHECK RESPONSIVENESS

CHECK PULSE

CHECK CIRCULATION

IF NO PULSE START CPR

IF NO CIRCULATION START CPR

IF NO PULSE CONTINUE CPR

IF NO CIRCULATION CONTINUE CPR

SHOCK ADVISED

TREATMENT ADVISED

NO SHOCK ADVISED

NO TREATMENT ADVISED

PRESS FLASHING SHOCK BUTTON

PRESS FLASHING TREATMENT BUTTON

RELEASE SHOCK BUTTON

RELEASE TREATMENT BUTTON

SHOCK DELIVERED

TREATMENT DELIVERED

NO SHOCK DELIVERED

NO TREATMENT DELIVERED