Qphy-sas2 test descriptions, Tx group 1 out of band tests 5.1.x (oob), Qphy-sas2 software option – Teledyne LeCroy QPHY-SAS2 User Manual
Page 27
QPHY-SAS2 Software Option
917718 Rev
A
27
QPHY-SAS2 TEST DESCRIPTIONS
Tx Group 1 Out of Band Tests 5.1.x (OOB)
There are 4 tests run in this group. The tests that are run are:
1. 5.1.2
– Tx Maximum Noise During OOB Idle
2. 5.1.3
– Tx OOB Burst Amplitude
3. 5.1.4
– Tx OOB Offset Delta
4. 5.1.5
– Tx OOB Common Mode Delta
After the completion of the Out of Band Tests the oscilloscope is in the following configuration:
Figure 17 - Oscilloscope Configuration after Out of Band Tests
Shown on this screen:
F3 is the Idle Differential Signal. This trace represents only the idle portion of the differential OOB signal.
F4 is the Burst Differential Signal. This trace represents only the burst portion of the differential OOB signal
after it is filtered by a 4.5GHz low-pass Butterworth filter.
Having these 2 traces separate from each other allows us to perform the measurement on the idle portion and the
burst portion separately.
F5 is the Burst Common Mode signal.
F6 is the Wide Band AM Demodulated Differential Signal. This is used for the separation of the idle portion
from the burst portion.
F7 is the Idle Common Mode signal. This is calculated by summing the two inputs and dividing them by 2.
F8 is a view of the Differential Signal after it has been filtered by the 4.5GHz low-pass Butterworth filter. This
is calculated by subtracting the negative input from the positive input. This signal is used to create many
of the traces described earlier.