Procedure, Figure 8 - lecroy sas protocol suite utility – Teledyne LeCroy QPHY-SAS2 User Manual
Page 16
16
917718 Rev
A
Procedure
1. Be sure the SAS Protocol Suite utility supplied with the SIERRA M6 is installed. (The directories below
may change depending upon your installation choices)
2. Copy the sas2 phy diagnostic page.sac file to the following directory:
C:\Users\Public\Documents\LeCroy\SAS Protocol Suite\User.
3. Copy the SAS2_PortA_Diagnostics.txt and SAS2_PortB_Diagnostics.txt files to the following director:
C:\Users\Public\Documents\LeCroy\SAS Protocol Suite\System\DataBlock.
4. Connect the SAS-2 device primary port output to the T1 front-panel connector of the Sierra. The primary
port is closest to the center of the SAS-2 multi-pin connector.
a. The LeCroy TF-SATA-C test fixture is required to connector the SMA connectors of the Wilder Tech
test fixture to the SATA connector on the front of the Sierra.
b. Test signals will be available on the secondary port.
5. Power on both the SAS-2 device and the SIERRA
6. Open the LeCroy SAS Protocol Suite utility.
Figure 8 - LeCroy SAS Protocol Suite Utility