Test equipment required, Operator’s manual – Teledyne LeCroy HFP2500 User Manual
Page 37

Operator’s Manual
922253-00 Rev A
31
Test Equipment Required
Table 3: Required Equipment lists the test equipment and accessories (or their equivalents) that are
required for complete calibration of the HFP2500 Probe. Specifications given for the test equipment
are the minimum necessary for accurate calibration. All test equipment is assumed to be correctly
calibrated and operating within the specification listed. Detailed operating instructions for the test
equipment are not given in this procedure. Refer to the test equipment manual if more information
is needed
If alternate test equipment is substituted, control settings or calibration equipment setups may
need to be altered.
Table 3: Required Equipment for Adjustment Procedure
Description Examples
Minimum Requirements
Example Test Equipment
Digital Oscilloscope
ProBus interface
Teledyne LeCroy WavePro960
Teledyne LeCroy LT344
Digital Multimeter (DMM)
with test probe leads
4.5 digit
DC: 0.1% Accuracy
AC: 0.1% Accuracy
Agilent Technologies 34401A
Fluke 8842A-09
Power Supply
0-12 V, settable to 10 mV
HP E3611A
BNC Coaxial Cable (2 ea.)
Male to Male, 50 Ω, 36" Cable
Pomona 2249-C-36 or
Pomona 5697-36
BNC Tee Connector
Male to Dual Female
Pomona 3285
Calibration Fixture
ProBus Extender Cable
Teledyne LeCroy PROBUS-CF01
Terminator, Precision, BNC
50 Ω ± 0.05%
Teledyne LeCroy TERM-CF01
Banana Plug Adapter (2 ea.)
Female BNC to Dual Banana Plug
Pomona 1269
BNC to Mini-grabber
BNC Male to Mini-grabber Cable,
36"
Pomona 5187-C-36