SKC 560-002 MethChek Immunoassay Wipe Kit User Manual
Page 6
6
5. With
pipett e, add 3 drops of extract liquid from
vial to the sample well (marked “S”) on Step 3
Cartridge.
6. Observe Step 3 Cartridge. Lines will appear
within one minute. Results will develop
completely within 5 minutes. See Reading
MethChek Results on page 7.
Do not disturb cartridge during sample
development.
7. Once test is complete, write ID on cartridge and
on mini bag, insert cartridge into mini bag, and
transport or store.
8. Place disposable items in resealable test packet
bag, seal, and dispose of as hazardous materials.
See Disposing on page 10.
Do not discard Step 1 Wett ing Agent solution,
spray nozzle, or bott le unless empty or
expired. Bott le contains enough solution for
several tests.
SKC recommends sealing the Step 1 solution bott le
with the provided screw cap for transport and
storage.
Thoroughly wash exterior of bott le with soap and water before
sampling at another site to avoid cross contamination.
Results appear in
≤1 minute. Development
completed in 5 minutes.
6
ID cartridge, transport or
store in mini bag.
7
Place disposable items in
resealable bag.
8
5
Pipette 3 drops of extract
into “S” well on Cartridge.