Preloaded Coated Filters

Cat. No. 225-9010

SKC coated fi lters are shipped preloaded in the cassett es with end plugs and
shrink bands in place. No assembly is required; they are ready to use.

Method & Chemical: OSHA 102 for acetic anhydride

Filter & Coating

: 2 glass fi ber fi lters coated with veratrylamine and di-n-octyl

phthalate. Four-piece 37-mm cassett e. Open-face sampling confi guration.

Prior to Sampling: Store at ≤ 39.2 F (4 C) for up to 3 months.

Sample Stability: Store samples at ≤ 39.2 F (4 C) or 71.6 F (22 C) for up to
15 days.

† Custom order due to very limited shelf-life.

For information on other available

coated fi lters, request SKC Publication 1219.

Operating Instructions

863 Valley View Road, Eighty Four, PA 15330

Tel: 724-941-9701 Fax: 724-941-1369

www.skcinc.com

How to Use the SKC Preloaded Coated Filters

1. Select one coated fi lter cassett e for calibrating the fl ow rate. A red plug
secures the inlet, which is clearly marked “inlet,” and a blue plug secures the
outlet. Set up the sampling train for calibration with the representative fi lter
cassett e in line. For “closed-face” sampling, remove the plugs and connect the
cassett e to the sampling train. For “open-face” sampling, remove the outlet
plug and the cassett e inlet section then connect the cassett e to the sampling
train. For details on sett ing up a sampling train, refer to the SKC Sample Setup
Guide “Sampling Train - Filters and Cyclones” (SKC Publication 1166).

2. Calibrate the pump to the desired fl ow rate using a fi lm fl owmeter or
calibrated precision rotameter. For details on calibrating a pump, refer to the
SKC Sample Setup Guide “Calibrating a Pump Using a Film Flowmeter” (SKC
Publication 1163).

3. Replace the cassett e used for calibration with a fresh coated fi lter cassett e
for sample collection. Att ach the cassett e to a worker’s collar, as close to the
breathing zone as possible, by using a fi lter cassett e holder (SKC Cat. No.
225-1). Sample for the specifi ed time interval and record the time.

4. Remove the fi lter cassett e at the end of the specifi ed sampling period and
replace both end plugs (and the inlet if necessary). Recheck the fl ow rate using
the same cassett e and fl owmeter used for calibration (in Step 2) to ensure that
the fl ow rate has not changed by more than 5%.

5. See Sample Stability on reverse side for sample storage. Appropriately package
filter cassettes and ship the samples, blanks, and all pertinent data to a
laboratory for analysis. Expedite shipping of samples to laboratory. Friday
shipments are not recommended.