SKC 225-9020 Preloaded Coated Filter User Manual

SKC Hardware

background image

Preloaded Coated Filters

Cat. No. 225-9020

SKC coated fi lters are shipped preloaded in cassettes with end plugs and shrink

bands in place. No assembly is required; they are ready to use.

Method & Chemical: OSHA 85 for n-valeraldehyde

Filter & Coating: 3 glass fi ber fi lters treated with 2,4-dinitrophenylhydrazine and

phosphoric acid. Three-piece 37-mm cassette. Open-face sampling confi guration

Prior to Sampling: Store at ≤ 39.2 F (4 C) for up to 3 months.

Sample Stability: Store samples in the dark at ≤ 39.2 F (4 C) or 71.6 F

(22 C) for up to 19 days. Refrigerated storage and shipment to the laboratory

is recommended.

For information on other available

coated fi lters, request SKC Publication 1219.

Operating Instructions

863 Valley View Road, Eighty Four, PA 15330

Tel: 724-941-9701 Fax: 724-941-1369

www.skcinc.com

SKC Limited Warranty and Return Policy

SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC's sole liability and the buyer's exclusive remedy. To view the complete
SKC Limited Warranty and Return Policy, go to http://www.skcinc.com/warranty.asp.

How to Use SKC Preloaded Coated Filters

1. Select one coated fi lter cassette for calibrating the fl ow rate. A red plug

secures the inlet, which is clearly marked “inlet,” and a blue plug secures the

outlet. Set up the sampling train for calibration with the representative fi lter

cassette in line. For “closed-face” sampling, remove the plugs and connect the

cassette to the sampling train. For “open-face” sampling, remove the outlet

plug and the cassette inlet section then connect the cassette to the sampling

train. For details on setting up a sampling train, refer to the SKC Sample Setup

Guide “Sampling Train - Filters and Cyclones” (SKC Publication 1166).
2. Calibrate the pump to the desired fl ow rate using a fi lm fl owmeter or

calibrated precision rotameter. For details on calibrating a pump, refer to the

SKC Sample Setup Guide “Calibrating a Pump Using a Film Flowmeter” (SKC

Publication 1163).
3. Replace the cassette used for calibration with a fresh coated fi lter cassette

for sample collection. Attach the cassette to a worker’s collar, as close to the

breathing zone as possible, by using a fi lter cassette holder (SKC Cat. No.

225-1). Sample for the specifi ed time interval and record the time.
4. Remove

the

fi lter cassette at the end of the specifi ed sampling period and

replace both end plugs (and the inlet if necessary). Recheck the fl ow rate using

the same cassette and fl owmeter used for calibration in Step 2 to ensure that

the fl ow rate has not changed by more than 5%.

5. See Sample Stability on reverse side for sample storage. Appropriately package
samples, blanks, and all pertinent data to send to a lab for analysis. Expedite
shipping of samples to laboratory. Friday shipments are not recommended.

Form 40049 Rev 1303