Sampling – SKC 690-101,-102,-103,-104 ULTRA III Passive Samplers User Manual

Page 2

background image

Sampling

1. Open pouch. Remove sampler.

a. Do not discard pouch.
b. Store pouch in a clean, organic solvent-free

environment.

2. Record the following:

a. Sample ID on pouch and sampler I.D. label
b. Sampling location on pouch
c. Date sample taken on pouch

Do not use Sharpie

®

markers or equivalent

to mark label or pouch as these can off-gas
VOCs.

3. Clip sampler(s) in appropriate location.

a. Indoors: Clip sampler(s) to stand accessory

(Cat. No. 690-302) and locate at appropriate
height.

b. Outdoors: Clip sampler(s) inside shelter

(open on two ends and bott om to allow air
to fl ow freely). Hang shelter securely at the
appropriate height.

Ensure diffusion holes in sampler are not
obstructed in any way.

4. Start sampling: Slide sampler cover to open

position (diff usion holes visible).

a. If performing for an extended sampling

time or sampling in an area of high
concentrations, install Secondary Diff usion
Barrier accessory (Cat. No. 690-300); see
Operating Instructions 39008
.

b. Write sample start time on pouch label.

5. Stop sampling: Slide sampler cover to closed

position (arrow and SKC visible).

a. Remove Secondary Diff usion Barrier, if used,

and then slide cover to closed position; see
Operating Instructions 39008
.

b. Write sample stop time on pouch label.

6. Place sampler in its resealable pouch

immediately aft er sampling.

Ensure sample ID on sampler matches sample
ID on pouch.

1

2

3a

3b

Sampler

I.D. label

4

5

6

Clip

Diffusion holes

(sample sorbent

compartment)

Sliding cover

Blank/

correction

sorbent

compartment

Clip

Sample compartment

Sliding cover

Blank/

correction

compartment