Aft er desired sampling period, unclip sampler
from worker's clothing. Immediately remove
green secondary diff usion barrier and dispose
following appropriate procedures.

Seal sampler with cover immediately.

It is important that samplers are sealed tightly

with cover. The samplers could potentially

pick up moisture over time which will affect the

recovery of N

2

O.

Write stop time on label on the pouch.

In a clean, organic solvent-free environment,
carefully remove plug from vial adapter on back
of sampler housing. Immediately screw open
end of sorbent vial into adapter (see inset 10A).
Rotate assembly until sorbent is draining from
sampler to vial. Tap sampler lightly. Ensure
all sorbent is recovered in the vial. Remove
sampler housing from vial and immediately
screw cap onto vial.

Avoid exposing sorbent to air.

Mark sample vial(s) as appropriate and seal in
aluminized pouch. Package sample(s) and one
blank sorbent vial in a padded envelope. Send
to a laboratory for analysis.

Storage

Before Sampling: Up to 6 months at < 39.2 F (4 C) in a clean environment

Aft er Sampling: 3 weeks at < 39.2 F (4 C) or 1 week at ambient temperature

in a clean environment

Sorbent vials should be capped tightly and kept in foil pouches before
and aft er use.