Performance Profi le

Sampling Rate

28.6 ml/min with a relative standard deviation of 7.6% at a wind

for Formaldehyde:

velocity of 5 to 100 cm/sec for 15 minutes to 24 hours

20.4 ml/min wind velocities < 5 cm/sec for 1 to 7 days

See the Passive Sampling Guide at www.skcinc.com for
UME

x

100 sampling rates for other aldehydes.

Validation Range:

0.06 to 3.0 ppm

Detection Principle:

Formation of stable DNPH-hydrazone in the

presence of formaldehyde*

Analysis Method:

Solvent extraction and analysis by HPLC (high-performance

liquid chromatography) with UV detection

Lower Detection Limits: 15 min: 200 ppb (.24 mg/m

3

)

8 hr: 5 ppb (.006 mg/m

3

)

24 hr: 2 ppb (.002 mg/m

3

)

7 days: 0.2 ppb (.0002 mg/m

3

)

Shelf-life:

12 mos from date of manufacture at

freezer temperatures (≤39.2 F [4 C])

Storage:

Before use: Freezer temperatures (≤39.2 F [4 C])

After use: Freezer temperatures (≤39.2 F [4 C]) and

analyze within 3 weeks

Capacity: 29

μg/sample

Accuracy:

± 25%, exceeds OSHA requirements

Temperature Effects:

No effect on sampling rate between 10 and 30 C

Humidity Effects:

No effect from 10 to 80% relative humidity (RH).

Do not use sampler below 10% RH.

Wind Velocity Effects:

No effect from 5 to 100 cm/sec

Interferences:*

Highly

specifi c for formaldehyde. Large amounts of

carbonyl compounds may reduce the uptake of formaldehyde

• Use in ozone levels less than 0.5 ppm

Dimensions:

3.4 x 1.1 x .35 in (8.6 x 2.8 x .89 cm)

Weight:

0.38 oz (10.9 gm)

Slide Cover: Green

* If sampling in an atmosphere containing formalin, see www.skcinc.com/instructions/1795.pdf for

fi eld study information.

Sampling Instructions

Cautions: • Store at freezer temperatures (≤ 39.2 F [4 C]) before use.

• Do not store with food.

• SKC recommends using gloves when handling chemically

treated media.

• Before sampling, check the expiration date on the pouch label.

Do not use aft er the last day of the month indicated.

UME

x

samplers are designed for single use only. Do NOT reuse

UME

x

samplers.

• Do not use Sharpie

®

markers or equivalent to mark label or

pouch as these can off -gas VOCs.