Calibration and Sampling

Note: As the particle load on the fi lter increases during sampling, the pressure
drop will also increase. Therefore, use a compensating sample pump such as
the Leland Legacy

®

(SKC Cat. No. 100-3002). The Leland Legacy pump will

provide 24 hours of run time at 10 L/min at a maximum pressure drop of
12 inches water.

Calibration

Calibrate pump fl ow rate with the IMPACT Sampler in line (loaded with fi lter
and impaction disc). See pump and calibrator operating instructions.

Ensure rain cover is removed

from inlet and that impactor is

loaded and fully assembled (see

Inserting a Collection Filter

into the IMPACT Sampler and

Inserting an Impaction Disc into

the IMPACT Sampler).

Screw Calibration Adapter onto

impactor inlet.

Using fl exible tubing, att ach pump inlet to outlet of impactor.

Use a short length of tubing to connect inlet of calibration adapter to outlet

of a calibrator to form a calibration train.

Calibrate the sample pump to 10 L/min.

See sample pump and calibrator

operating instructions.

3

4

5

1

When calibration is completed,

disconnect calibrator and tubing

from adapter. Remove calibration

adapter from impactor.

6