Media and Impactor Preparation

Equilibrate and pre-weigh the fi lters in a clean environment according to appropriate
procedures. Record the weight as the pre-sample weight.

Clean the Sioutas Impactor in between samples to remove any residual particles.
See Cleaning.

Assembly

In a clean environment, assemble the Sioutas Impactor from outlet to inlet as
follows:

Caution: Handle all fi lter material with forceps.

Caution: Ensure that the screen and spacer ring (Figure 2) are in place before
inserting fi lter. The screen prevents the fi lter from restricting the outlet.

5

6

7

Repeat Steps 3 and 4 for stages B-D.

Remove O-ring from outlet plate

(see Figure 2).

Remove compression ring from outlet

plate (see Figure 2).

1

Remove O-ring and compression ring

from outlet plate (see Disassembly,

Steps 6 and 7 and Figure 2).

Insert 37-mm aft er-fi lter in outlet plate.

Replace compression ring and O-ring.

Figure 2. Outlet plate, exploded view

O-ring

Compression

ring

Spacer ring

Screen

Outlet

plate

Note: Use forceps to carefully insert or remove impaction substrate, aft er-fi lter,