Cassette Holder (Figure 2)

1. Insert the cyclone body through

the large opening of the cassett e
holder.

2. Ensure the cyclone/cassett e

assembly is seated fi rmly in the
holder by inserting the cyclone’s
small round nodule on the rim
into the notch in the cassett e
holder.

3. Remove the plug from the

cassett e outlet.

4. Secure the cyclone/cassett e

assembly in the holder by
stretching the spring-loaded
hold-down plate over the
cassett e outlet.

5. Insert the holder’s Luer adapter

(located on the end of the rubber
tubing) into the cassett e outlet.

Calibration

Option 1: Using a Calibration Jar (Figure 3)
1. Prepare a cyclone/cassett e assembly (see Figure 1). Ensure grit pot remains

on cyclone body during calibration.

SKC recommends using the smallest calibration jar possible. To

achieve this, do NOT use Cassette Holder 225-1 during calibration.

2. Place cyclone/cassett e assembly into an airtight calibration jar that contains an

inlet and outlet (SKC Cat. No. 225-111, see Figure 3).

3. Using fl exible tubing, connect inlet of calibrator to inlet of calibration jar.
4. Run tubing from outlet of cyclone/cassett e assembly through outlet of

calibration jar and to the inlet of a sample pump. If using the SKC 225-111 jar,
connect the Luer adapter inside the jar to the cyclone/cassett e outlet. Connect
jar outlet to inlet of a sample pump.

Figure 3

Tygon tubing

Rubber tubing

Filter cassette

Filter cassette
holder

Grit pot

Collar clip

Luer adapter

GS cyclone

Spring-loaded
hold-down plate

Cassette outlet

Figure 2

GS-3 Cyclone/Cassette Assembly

in Holder (Cat. No. 225-1)

Calibrator

GS-3 Cyclone/cassette

assembly inside calibration jar

Inlet

Inlet

Sample pump

Outlet

Multi-purpose
calibration jar
(225-111)