SKC 224-PCXR4 Universal Pump User Manual
Page 5

3
RFI/EMI Shielding:
Complies with requirements of EN 55022, FCC Part 15 Class
B, EN 50082-1; frequency range of the radiated susceptibility test
was 27 to 1000 MHz.
Flow Fault:
If the pump is unable to compensate for longer than 15 seconds due
to excessive back pressure, the pump enters fl ow fault. During fl ow
fault, the pump stops, displays FLOW FAULT, pauses timing func-
tions, and displays elapsed time. Auto-restart is attempted up to 5
times.
Low Battery Fault:
Pump stops, displays LO BATT, pauses timing functions, and dis-
plays elapsed time.
Battery Test:
LCD shows battery condition prior to sampling.
Time Display:
LCD displays up to 9999 minutes (6.8 days) at which point the
display rolls over to 0.
Timing Accuracy:
± 0.05% (± 45 seconds per day)
Sampling Pause (Hold): Allows user to temporarily halt sampling without loss of timing data.
Restart does not require resetting time.
Tubing:
Requires 1/4-inch ID tubing
CE marked
UL Listed
ATEX-approved models available