Ronan X90 Series User Manual
Page 6

module is not necessary in order to change wiring. 
Grommeted holes are provided on the top and 
bottom surfaces of the chassis for cable entry and 
exit. Overall dimensions are 7.25” high X 4.5” wide 
X 5.75” deep (18.42 cm X 11.43 cm X 14.6 cm). 
 
Refer to drawing number C-1359-K for terminal 
designations for power wiring and hook-up to the 
remote probe driver board. 
 
7.3 Integral Model X90-1005 
 
Refer to drawing A-1697-K for rear-panel input 
power wiring connections and relay contact wiring 
connections. 
 
————————————————————————— 
 
8.0 CALIBRATION CONTROLS
 
The following front-panel controls are used in the set-up 
procedure. 
 
ZERO CONTROL is used to provide a zero voltage 
signal at TP-AMP with the process level below the point 
of measurement, under conditions of maximum radiation 
at the G-M tube. 
 
SPAN is used to provide a + 1.0 voltage signal at TP-
AMP when the process level is above the point of 
interest, under conditions of no radiation on the G-M 
tube. 
 
Three light-emitting diode (LED) DISPLAYS are 
provided. These indicators will signal normal and level 
alarm for process conditions and failure alarm. 
 
The AMP TERMINAL is the output of the voltage 
integrator. This output increases as the radiation on the 
tube decreases and vice versa. 
 
Calibration should be attempted only after all system 
components have been securely mounted in position and 
power and other connections made where required. 
Check also that the correct jumpers have been installed 
on the X90-301(V) or X90-1005 printed circuit board. 
 
————————————————————————— 
 
9.0 CALIBRATION—GENERAL
 
The output at the AMP terminal is a measure of level 
present at the detector. This reading will be minimum or 
near zero when the detector is fully irradiated (no 
process) and the maximum or near 1.0 V when there is 
process material between the source and the detector. 
 
Therefore, for calibration any option, two voltage 
readings are generated: 0.0 V with no process between 
the source and the detector and + 1.0 V with process 
between the source and the detector. 
 
 
9.1 Calibration with Process Material
a) Switch on the power to the X90 and turn the 
shutter handle on the source holder to the ON 
position. Turn the SPAN control clockwise 20 
turns; then 10 turns counterclockwise. 
 
b) Lower the process level below the set limit and 
adjust the ZERO control so that the AMP terminal 
output is zero. 
 
c) Raise the process level above the set limit and 
again measure the voltage at the AMP terminal. 
 
d) Adjust the SPAN control to obtain a reading of + 
1.0 V at the AMP terminal. 
 
After the switch has been calibrated using the 
above method, the operation of the control relay, 
NORMAL and ALARM indicators should be 
checked by raising and lowering the process level 
about the set limit. If all these operate as required, 
the switch is considered calibrated. 
 
9.2 Calibration without Process Material
a) Turn the source holder shutter to the ON 
position. 
 
b) Turn the SPAN control clockwise 20 turns; then 
10 turns counterclockwise. 
 
c) Take voltage reading at the AMP terminal. 
Adjust the zero control to obtain an average 0.0 V 
reading at the AMP terminal. 
 
d) Turn the source OFF. 
 
e) Adjust the SPAN control for an average + 1.0 V 
reading. 
 
f) Turn the source back ON for operation. 
 
9.3 Time Delay
If false alarms occur, increase the time delay by 
moving the delay jumpers on the printed circuit 
board to a higher value. 
 
9.4 Response Time 
The response time of the amplifier is 0.5 to 40 
seconds, depending on the SPAN setting. 
Device
Install
High limit switch
High and G
Low limit switch
Low and G
Process alarm only
A-10, B-11, C-12
Process alarm and failure 
alarm
Failure relay and D-10, E-11, 
F-12
Delays
Select one: 0.1 sec., 5 sec., 
10 sec., 20 sec.
