Full specifications – General Tools and Instruments IRT659K User Manual
Page 11
11
2) Press the IR/K (TC) button (Fig. 1, Callout I). Note that this causes the
IRT LOG information on the bottom line of the LCD to disappear and
disables the MODE/SET and
ĵ
LOG buttons.
3) Attach the end of the probe to the surface whose temperature you wish to
measure. The measured value will appear on the main readout.
If you allow the IRT659K’s 30-second Auto Power Off function to activate with
the unit in thermocouple measurement mode, when you power the unit back on
it will enter IRT measurement mode.
FULL SPECIFICATIONS
Distance-to-Spot (D:S) Ratio
12:1
IR Temperature Measurement Range
-58° to 1202°F (-50° to 650°C)
IR Temperature Measurement Accuracy
±3°F (2°C) or ±2% of reading (whichever is
greater) above 32°F (0°C); ±4°F (3°C) or
±3% of reading (whichever is greater)
below 32°F
“K” Port Measurement Range
-146° to 2498°F (-99° to 1370°C)
“K” Port Measurement Accuracy
±(1% of reading + 1°C) above 0°C;
±(2% of reading + 2°C) below 0°C
Included “K” Type Thermocouple
0° to 500°F (-18° to 260°C)
Measurement Range
Display Resolution (IR or TC Measurement) 0.1° (F or C)
Emissivity
Adjustable from 0.1 to 1.0 in 0.01 steps,
with default of 0.95
Automatic Measurement Storage Capacity Last 10 readings
Display Size & Type
1.5 in. (38mm) diagonal backlit LCD
Response Time
500 ms
Laser Pointer Wavelength
655nm
Laser Power
Class 3R (<3mW)
Current Consumption
30mA, max
Operating Temperature
32° to 104°F (0° to 40°C) @ 10 to 75% RH,
non-condensing
Storage Temperature
-4° to 140°F (-20° to 60°C)
Auto Power Off Trigger
30 seconds of inactivity