4. PH CALIBRATION PROCEDURE

4-1 Calibration Consideration

The most ideal PH ELECTRODE generates 0 mV at PH
7.00 ( 177.4 mV at PH 4 ) and YK-2001PH has been
always calibrated with signals which simulate the most ideal
PH ELECTRODE ( based on 25
ambient environment
). However not every PH ELECTRODE is as accurate as
the most ideal one, so calibration procedures are
necessary to be done before the first time measurement.
In addition to the first time measurement, users are also
recommended to execute the calibration procedures to
ensure the high accuracy measurement.

4-2 Required Equipments for Calibration

1) PH ELECTRODE ( optional ).

2) PH buffer solutions ( optional ).

4-3 Two Points Calibration Procedure

1) Power on the instrument by pressing the " Power

Button " ( 3-2, Fig. 1 ).

2) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to let meter be

operated under the PH function with a " PH " symbol on the

display.

3) Adjust the " Temperature Compensation Value " to make

it same as the temperature value of the pH buffer

solution.

Manual temperature compensation value adjustment

procedure, please refer to 5-1 (Page 10).

Automatic temperature compensation value adjustment

procedure, please refer to 5-2 (Page 14).

7