4-5 Others

Above calibration procedures effect only when reading value

within 1 PH of the calibration point.

However if the reading value beyonds

* 1 PH of PH 7 ( > PH8, < PH6 )

* 1 PH of PH 4 ( > PH5, < PH3 )

* 1 PH of PH 10 ( > PH11, < PH9 )

the calibration procedures are :

1) Connect the PH ELECTRODE to the " PH BNC Input

Socket " ( 3-10, Fig. 1 ).

2) Power on the instrument by pressing the " Power

Button " ( 3-2, Fig. 1 ).

3) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select

the PH function with a "PH" symbol on the display

4) Set the " Manual temperature compensation " value to 25 蚓

refer to 5-1 calibration procedures, page 11.

5) Place the electrode into the standard solution ( PH7, PH4 or

PH10 ) , then the instrument will have the PH value on the

display.

6) * For measuring the PH7 standard solution, adjust the

PH 7 VR ( 3-15, Fig. 3 ) until the display value

within PH6 to PH8.

* For measuring the PH4 standard solution, adjust the

PH 4 VR ( 3-16, Fig. 3 ) until the display value

within PH3 to PH5.

* For measuring the PH10 standard solution, adjust the

PH 10 VR ( 3-16, Fig. 3 ) until the display value

within PH9 to PH11.

7) The following calibration procedures will be same as 4-3

( page 7 ) and 4-4 ( page 9 ).