General Tools and Instruments DPHC2001 User Manual
Page 15
@
Use the " Up Button " ( 3-5, Fig. 1 ) and the " Down
Button " ( 3-8, Fig. 1 ) to adjust the PH 4 default
calibration value.
@
The adjustment range of " PH4 default calibration value "
is limited within 4.0 0.20 PH
c. Adjust the PH 7 default value
*
When the PH 4 default value be adjusted. Press
the " TEMP. C Button " ( 3-7, Fig. 1 ) once
again to adjust the PH 7 default calibration
value.
The upper display will show the values of " 7.00 " and
the lower display will show the " PH 7 default
calibration value ".
PH
7.00
07.12
@
Use the " Up Button " ( 3-5, Fig. 1 ) and the "
Down Button " ( 3-8, Fig. 1 ) to adjust the PH 7
default calibration value.
@
The adjustment range of " PH 7 default calibration
value " is limited within 7.0 0.20 PH.
d. Adjust the PH 10 Default Value
*
When the PH 7 default calibration value be
adjusted. Press the " TEMP. C Button " ( 3-7,
Fig. 1 ) once again to adjust the PH 10 default
calibration value.
13
- 119 (1 page)
- 146 (2 pages)
- 1478 (1 page)
- 147 (1 page)
- 80560 (1 page)
- 840 Pro Doweling Kit (36 pages)
- 840 use of Dowel Centers (4 pages)
- 841 (8 pages)
- 849 (2 pages)
- 850 (2 pages)
- 860 v.1 (16 pages)
- 860 v.2 (44 pages)
- 860 Addendum (1 page)
- 870 v.1 (2 pages)
- 870 v.2 (46 pages)
- 880 (2 pages)
- 861 (16 pages)
- AQ150 (16 pages)
- AT60LR (1 page)
- BAR4225 (13 pages)
- BF10 (12 pages)
- CA10 (16 pages)
- CAF4221 (12 pages)
- CAF4224 (14 pages)
- CDM77232 (15 pages)
- CDM77535 (12 pages)
- CGD900 (12 pages)
- CIH20DL (28 pages)
- CL10 (20 pages)
- CMM880 (11 pages)
- CMR35 (36 pages)
- CPH12101 (1 page)
- CT101 (8 pages)
- CT102 (8 pages)
- CT103 (8 pages)
- CT6235B (16 pages)
- DA833 (16 pages)
- DAF2005MDL (27 pages)
- DAF3300 (40 pages)
- DAF3010B (12 pages)
- DAF4207SD (16 pages)
- DAF4223 (13 pages)
- DAF80PWM (10 pages)
- DAF80PW (10 pages)
- DBAR110 (16 pages)