The weigh rail, its approaches, the load cells, and their housing are
arranged in a configuration that requires no space above the customer's
normal supporting structure. All components of the Unirail are
assembled as one complete calibrated unit ready for installation into the
existing monorail line. Install the Unirail at the approach rails by direct
attachment to the rail hangers. The scale should be installed level in
both directions in the horizontal plane. If level condition along the rail
cannot be obtained, out of level must not exceed 0.6 degrees, for 0.5
feet on the length of the weigh rail. There is no welding required to
install the scale. If any welding is required on any adjacent structure,
perform the welding prior to scale installation. If welding is required after
installation, the entire scale assembly, live and adjacent dead rails,
should be physically removed prior to welding to avoid possible damage
to the load cells.
W
W
ARNING:
ARNING:
DO NOT weld on the scale structure or damage to the load cells
DO NOT weld on the scale structure or damage to the load cells
may occur
may occur
.
.
Note:
Use cast or fabricated steel hangers to support the scale structure.
Do not use light gauge (cold formed) steel hangers.