Section 5: Troubleshooting Guide

Error code on
LCD display

Description

Error

Error in calibration procedure.

Err

Weighing function disabled. Scale is unstable or error in application.

Err.OL

Overload or Underload condition

L.bAtt Low

battery.

SLEEP

Instrument is in the sleep mode.

“------”

Zero key pressed and instrument is unstable and unable to weigh.




Section 6: Rechargeable Battery

Battery life
continuous

Recharge time

100

8 to 12 hours

NOTES:

There are four (4) basic tests for a battery using a digital volt meter and the
front panel indications.

1. Any battery not connected to an instrument or charger and having a
reading less than 4.5 VDC should be disposed of properly and replaced.

2. A fully-charged battery not connected to an instrument or charger
should have a reading of 6.4 VDC.

3. A fully-charged battery plugged into and loaded by an instrument (AC
adapter unplugged) should have a reading of 6.25 to 6.60 VDC as seen
from the front panel.

4. A fully-charged battery plugged into and loaded by an instrument (AC
adapter plugged in) should have a reading of 9.60 VDC as seen from the
front panel.