Section 3: Scale Installation

08/12

15

51102 Rev. 7

If corners

do

require adjustment, complete the following steps:

1. Place the concentrated weight on the corner displaying the lowest weight.

2. Turn the adjustment on the potentiometer clockwise (

CW

) to the displayed weight

so it reads the same as the highest reading.

3. Repeat this procedure while rechecking all corners until they are equal.

Important Note:

When moving the weight(s) from corner to corner,

DO NOT

zero the scale. The purpose is to adjust the corners to be the same, and not to
perform a correct calibration.

4. Perform a zero reference check with an unloaded platform,

5. Repeat the corner test to ensure all readings are the same before proceeding.

If corners

do not

require adjustment, complete the following steps:

1. Remove all weights.

2. Zero the indicator.

3. Perform a final calibration with test weights.

4. Follow the appropriate indicator service manual to ensure a proper calibration.