Calibration steps – Fairbanks Aegis Industrial Mild Steel User Manual

Page 14

background image

Section 3: Scale Installation

08/12

14

51102 Rev. 7

Calibration Steps

Adjust the analog interface indicator to the platform.

Install all the corners to within

one (1) division of each other at 25% of

rated capacity.

Follow the appropriate indicator service manual to ensure a proper calibration.

STEPS

1. Center the four Junction Box Potentiometers by turning the adjustment screw

counter-clock-wise position until a clicking sound is heard, then turning each of
them back clock-wise

ten (10) turns

.

Total number of turns is twenty.

2. Identify the platform corner numbers.

3. Place a concentrated weight (

25%

of platform capacity)

on corner #1, then move it to #2, #3 and #4, noting the
displayed reading on each corner.

4. Identify the lowest reading, and then place the

concentrated weight on this corner.

1

2

3

4

50612-1