Teledyne 9070 - O2 and CO2 Food package analyzer (preliminary) User Manual

Page 8

background image

Outputs
• Two isolated linearised 4–20 mA DC outputs into 1000Ω load (max).
• RS232/485 computer/printer interface for peak oxygen/carbon dioxide value report and alarm functions.
• One common alarm relay for self diagnostic alarms
• One user selectable alarm relays for gas related alarm levels, ‘Sensor low temperature’ and ‘Calibration in

progress’.


Power
• 240/110 VAC, 50/60 Hz, 115W

Gas Connection
• 1/8’’ Swagelok tube connector

Flow Rate
• 200 - 300 cc (25-40 scfm) per minute, governed by internal pump

Environmental
• 0–50°C (32-120°F) ambient temperature
• 0–45°C (32-110°F) ambient temperature when CO2 sensor fitted


Weight
• 4.5 kg (10 lbs)

Dimensions
• 265mm x 158mm x 268mm. (9.91” x 6.22” x 10.55”)

Range of Output 1
• Field selectable from the following:

Output Selection

Range

Linear oxygen

0–0.1% oxygen to 0–100.0% oxygen

0–1000ppm oxygen to 0–1,000,000ppm oxygen

Low Range Linear oxygen

0–0.001% oxygen to 0–1.0% oxygen

0–10ppm oxygen to 0–10,000ppm oxygen


Range of Output 2
• Field selectable from the following:

Output Selection

Zero Range

Span Range

• Carbon Dioxide,

0–90%

10–100%

Min span 10%

• Reducing Oxygen

10-1–10-10 %

10-1–10-30 %

In one decade steps. Min span three decades

• Oxygen sensor EMF

0–1100 mV

1000–1300 mV

In 100 mV steps

In 100 mV steps

• Sample Oxygen

0–0.1%

0.01–20%

Min Span 0.01%

0–1000 ppm

100–20,000 ppm

Min span 100 ppm

• Low oxygen

0–99.9%

0.1–100% oxygen

Min span 0.1%

0-999,000 ppm

1,000-1,000,000 ppm

Min span 1,000ppm

• Logarithmic Oxygen

0.1% O2 Fixed

20% O2 Fixed


Range of Indication, Upper Line
• Oxygen selectable either % O2 or ppm, and Carbon Dioxide

• Oxygen, auto ranging from 0.1 ppm to 100% O2 (always ppm below 0.1% oxygen if selected as % O2 in set-up

#27)

• Carbon Dioxide, 0.1 to 100.0%

Page 8

Teledyne Analytical Instruments 9070 Oxygen Analyzer