Teledyne 9070 - O2 and CO2 Food package analyzer (preliminary) User Manual

Page 34

background image

1. Linear oxygen (0.1 to 100.0%)

*

2. Low range linear oxygen (10 to 10,000 ppm)
The output spans are adjustable in Set-up steps 21 and 22.

21. TRANSMITTER ZERO CHANNEL 1

Select transmitter zero for output Channel 1.
Range 0.0 to 99.9% Oxygen.
Range 0 to 0.999% (9,990ppm) Oxygen.
Default setting is 0.0%.

22. TRANSMITTER SPAN CHANNEL

Select transmitter span for output Channel 1.
Linear Oxygen-

Range 0.1 to 100.0% Oxygen. Default setting is 10.0%.

Low Oxygen-

Range 0.001 to 1.000% (10 to 10,000ppm) Oxygen. Default setting is 0.100%.

23. TRANSMITTER OUTPUT CHANNEL 2

Select the type of output required for Channel 2.
Options:
1. Sample Gas, oxygen
2. Linear (Low) Oxygen, 0.1 to 100.0 % *
3. Logarithmic oxygen, 0.1 to 20 %
4. % carbon dioxide
5. Reducing oxygen %
6. Oxygen sensor EMF

24. TRANSMITTER ZERO CHANNEL 2

The output zero and span of Channel 2 is set in set-up steps 24 and 25. Range limits are shown below.

25. TRANSMITTER SPAN CHANNEL 2

Output

Zero Range

Span Range

Minium span

SAMPLE GAS, OXYGEN

0.0 to 99.9 %

0.1 to 100 %

0.1 %


LOW OXYGEN

0.0 to 99.9 %

0.1 to 100 %

0.1 %


LOG OXYGEN

0.1 % oxygen fixed

20 % oxygen fixed

(see Note 1)


CARBON DIOXIDE

0 to 80 %

20 to 100.0 %

20 %


REDUCING OXYGEN

10

-

1

to 10

-28

%

10

-

2

to 10

-

30

% Two

decades

(see Note 2)

oxygen in one

oxygen in one

decade steps

decade steps.

SENSOR EMF

0 to 1100 mV in

100 to 1300 mV

100 mV

100 mV steps

in 100 mV steps


NOTE
1:

For log oxygen scale details, Refer to Appendix 2.

2:

Note that the reducing oxygen span is shown on the display as the exponent only. -1 represents 10

-1

% oxygen.



26. SAMPLE MODE

Options
1. Fast Sample
2. Display Sample

*

3. Continuous

Page 34

Teledyne Analytical Instruments 9070 Oxygen Analyzer