Teledyne 9070 - O2 and CO2 Food package analyzer (preliminary) User Manual

Page 31

background image

Setup

Alarm

Display

Purge

Run

Enter

Option

Function

Option

Function

Autocal

Cal 2

Cal 1

Power

Alarm

Setup

Oxygen % 2.83
SensorTem p 718C

9070 Oxygen Analyzer Keyboard


5.5

SET-UP FUNCTION DETAILS

Note: The * indicates the default setting after a COLD-START. See Section 6.1

1. Calender year

Options
Select the current year for the real time clock/calender.
The cold start default sets the date and time to the software version date.

2. Calender month

Options
Select the current month for the real time clock/calender.

3. Calender day

Options
Select the current day for the real time clock/calender.

4. Real time clock hour

Options Select the current hour for the real time clock. (24 hour format)

5. Real time clock minutes

Options
Select the current minutes for the real time clock.

6. Reference voltage # 1 ( about 27.5 mV’s )

Options
Set the value of the reference voltage as read on a 3 3/4 digit multimeter (See Section 6.2 for further details).
27.55 mV

*

7. Reference voltage # 2 ( about 194 mV’s )

Options
Set the value of the reference voltage as read on a 3 3/4 digit multimeter (See Section 6.2 for further details).
193.60 mV

*

8. Reference voltage # 3 ( about 1200 mV’s )

Options
Set the value of the reference voltage as read on a 3 3/4 digit multimeter (See Section 6.2 for further details).
1202.00 mV

*

Teledyne Analytical Instruments 9070 Oxygen Analyzer

Page 31