3.0 Installation

Teledyne Analytical Instruments

A Business Unit of Teledyne Electronic Technologies

sis. TAI recommends that a bypass filter assembly in the sample loop be
installed.

Figuire 3-1.

Sampling Considerations

3.2.2 Effluent Return

Returning the sample effluent to the process may require a separate

calibration fluid return so that the calibration fluids can be kept out of the
process during standardization periods.

3.2.3 Flow Control

A differential pressure regulator is often used in TAI systems across the

sample flowmeter and needle valve control to give a constant flow, regard-
less of upstream or downstream pressure fluctuations. Constant flow may or
may not be important, depending upon the application.

It is necessary to insure sufficient contact of the sample with the

preheater in order to bring the sample up to the control temperature. Do not
allow flow in excess of 50 ml/minute.

For liquid applications, TAI recommends that the throttle valve and

flowmeter always be downstream from the analyzer. Such an installation
will pressurize the sample cell and eliminate bubbles in the sample.
Bubbles suspended in the sample fluid will produce erratic, ambiguous
analysis of the sample.

Analyzer Flow Cell

Sample Inlet

Regulator
Bypass Valve

Sample
Flowmeter

Diff. Regulator

Ring Manifold

Bypass
Flowmeter

Sample Outlet

Filter installed only
when required

Sample

Zero Fluid

Span Fluid

Cell Drain

Calibration
Fluid Return

Sample Return