2.0 Operational Theory

Teledyne Analytical Instruments

A Business Unit of Teledyne Electronic Technologies

Figure 2-2

Analyzer System - Block Diagram

2.1.3 Power Module

See Figure 2-3. The power module controls power to the analyzer unit,

providing the switching function for the local meter, and providing tempera-
ture control for the sample and detector modules. In the case of the explo-
sion-proof configuration, where the control unit is mounted locally, the
power module simply routes the AC input power to its destination and
allows the control unit to provide the ON/OFF function.

When power is applied to the system, it is directed to the constant

voltage transformer and to the three temperature controllers which are
insensitive to line voltage fluctuations.

In order to facilitate easier calibration and to provide a quick visual

indication of the instrument’s status, a local meter is provided. With the