beautypg.com

Operational theory 2.0, Figure 2-5. control module - block diagram – Teledyne 514 - NDIR analyzer User Manual

Page 23

background image

2–11

Operational Theory 2.0

Teledyne Analytical Instruments

A Business Unit of Teledyne Electronic Technologies

Power for the control module is provided by a center-tapped trans-

former which takes the 115 VAC input, reduces it to 40 VAC, and feeds the
voltage to a DC power supply identical to the one installed in the detector
module. Power supply outputs are +24 VDC unregulated, and +15 VDC
regulated.

Ext.

Volt.

Amplifier

Alarm

Comparator

Auto

Zero

Timer

Manual

Setpoint

Adjust.

#1

#2

Relays

N.O.

Com.

N.C.

N.O.

Com.

N.C.

E-to-I

Converter

Current

(4

to

20

mA)

Standard

Valve

Control

Signal

Span

Adjust.

Meter

Indication

Voltage

(0

to

1

V

f.s.)

Millivolts

(0

to

5

mV

f.s.)

To

I-to-E

Converter

To Timer

15 VDC

24 VDC

Power

Supply

Power

Transformer

Manual

Fine

Zero

Power

115

VAC

60

Hz

Input

I-to-E

Converter

Current

Signal

Input

From

Analysis

Unit

115

VAC

60

Hz

Output

±

±

Figure 2-5.

Control Module - Block Diagram