background image

Total Hydrocarbon Analyzer

Operation

Teledyne

Analytical

Instruments

37

4.6.4.2

S

ETUP

M

ODE

The MAIN MENU consists of functions you can use to customize

and check the operation of the analyzer. Figure 4-3 shows the functions
available in the Model 4020. They are listed here with brief descriptions:

AUTO-CAL:

(Option) Used to define and/or start an

automatic calibration sequence.

PSWD:

Used to establish password protection or change the

existing password.

LOGOUT:

Logging out prevents unauthorized tampering with

the analyzer settings.

MODEL:

Displays Manufacturer, Model, and Software

version of the instrument.

SELF-TEST:

The instrument performs a self-diagnostic

routine to check the integrity of the power supply, output
boards, cell and amplifiers.

SPAN:

Set up and/or start a span calibration.

ZERO:

start a zero calibration.

ALARMS:

Used to set the alarm setpoints and determine

whether each alarm will be active or defeated, HI or LO
active, and failsafe or not.

RANGE:

Used to set up three analysis ranges that can be

switched automatically with auto-ranging or used as
individual fixed ranges. A digital range selection mode is
incorporated that uses the remote calibration inputs for range
selection.

 FILTER: This function sets the digital filter from 0 to 7, 0

being the fastest response time, and 7 being the slowest. The
default is 4.

 ANALOG-OUT ADJUST: Calibrate Analog Output
 LINEARIZATION: A restricted function not generally

accessed by the end user. It is used to linearize the output for
the range of interest.