Installation Model

4020

Teledyne

Analytical

Instruments

20

Table 3-3: Analog Range ID Output—Example

Range

Voltage Signal

Current Signal

Application

Output (VDC)

Output (mA DC)

Range 1

0.20

6.8

0–1 ppm CH

4

Range 2

0.50

12

0–10 ppm CH

4

Range 3

0.80

16.8

0–100 ppm CH

4

3.3.3.5

A

LARM

R

ELAYS

The nine alarm-circuit connector pins connect to the internal alarm

relay contacts. Each set of three pins provides one set of Form C relay
contacts. Each relay has both normally open and normally closed contact
connections. The contact connections are shown in Table 3-4. They are
capable of switching up to 3 amperes at 250 VAC into a resistive load.
The connectors are:

Threshold Alarm 1:

 Can be configured as high (actuates when

concentration is above threshold), or low (actuates
when concentration is below threshold).

 Can be Configured as failsafe or non-failsafe
 Can be configured out (defeated).

Threshold Alarm 2:

 Can be configured as high (actuates when

concentration is above threshold), or low (actuates
when concentration is below threshold).

 Configured as failsafe or non-failsafe
 Can be configured out (defeated).

System Alarm:

 Actuates when DC power supplied to circuits is

unacceptable in one or more parameters. Permanently
configured as failsafe and latching. Cannot be defeated.