pH
pH 2700, ION 2700, PC2700 only
Range
-2.000 to 20.000 pH
Resolution
0.1 / 0.01 / 0.001 pH
Accuracy
±0.002 pH + 1 LSD
Cal. Points
Up to 6 preset or 5 custom
Buffer Sets
USA, NIST, DIN, User1, User2, Custom
Slope Display
Yes (with offset)
Temp Compensation
Automatic or Manual (0 to 100 ºC / 32 to 212 ºF)
Temp Range (Meter)
0.0 to 100.0 ºC / 32.0 to 212.0 ºF
Inputs
BNC, ATC, Reference (Half-cell)
mV
pH 2700, ION 2700, PC2700 only
Range
±2000.0
mV
Rel. mV Range
±2000.0 mV
Resolution
0.1
mV
Accuracy
±0.2
mV
Offset Adjustment
Up to ±150 mV
Ion
ION 2700, PC2700 only
Range
0.001 to 19999 ppm (±2000 mV)
Resolution
2 or 3 digits
Accuracy
0.5% full scale (monovalent)
1% full scale (divalent)
Cal Points
2 to 8 points;
0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000
Slope Range
15 to 90 mV/decade
Temperature
ALL MODELS
Resolution
0.1 ºC / 0.1 ºF
Accuracy
±0.3 ºC / ±0.5 ºF
Calibration
Offset in 0.1 º increments;
Offset range: ± 5 ºC / 9 ºF