Nova-Tech 2700 Series Benchtop Meters User Manual
Page 34

30
Before and after a two point ion calibration using 0.001 and 0.01 standards
5) The mV/decade slope value will be displayed if the calibration is successful.
“Slope Error” indicates that the calibration for the current point was not
successful. This occurs when the slope (mV difference between two
consecutive points) is < 15 mV/decade or > than 90 mV/decade.
6) Repeat steps 4 & 5 with additional ISE standards or press ESC to escape
and save your calibration.
At left; a three point ion calibration, at right; the Ion calibration report
To view the Ion Calibration Report at any time from the measure screen, press
REPORT.
See also other documents in the category Nova-Tech Equipment:
- B4000-16 (28 pages)
- B4000-M (23 pages)
- Foxx Life Sciences (2 pages)
- CON 700 (24 pages)
- pH-CON 700 (32 pages)
- Oakton Testr 35 (6 pages)
- 4066, 4067, 4068, 4069, 4161, 4162, 4173, 4174, 4176, 4177 Control Company (4 pages)
- 4065, 4175, 4270, 4271, 4274, 4565, 4570, 4574 Control Company (4 pages)
- 4170 Control Company (4 pages)
- WD-35653-15 (3 pages)
- WD-35653-13 (1 page)
- pH-CON 300 (26 pages)
- pH-CON 10 (29 pages)
- CON 6+ (40 pages)
- 4070 Control Company (1 page)
- WD-00653-00 Zero Oxygen (1 page)
- pH-Dissolved Oxygen Meter (35 pages)
- DO 700 (36 pages)
- DO 6+ (48 pages)
- Foxx Autofil Bottle Top Filtration (2 pages)
- Foxx Autofil Centrifuge Funnel and Stand (2 pages)
- Foxx Autofil Laboratory Storage Bottles (2 pages)
- Foxx Autofil Pedestal and Cradle Ring (4 pages)
- RITEFLOW FLOWMETERS AND KITS (16 pages)
- Roto-Flo Flow Indicators 19937 Series (1 page)
- Leveling Base Kit (each) H40404-1000 (1 page)
- Taylor M-3000 TTi3000 Colorimeter Series (28 pages)
- Bel-Art Magnetic Stirring Bars Selection & Use Guide (4 pages)
- VISPOL-Series Polarimeters (2 pages)
- UniPol L-Series Polarimeters (4 pages)
- iPR-Series (8 pages)
- Atago AP-300 Series Polarimeter (4 pages)
- Foxx Heavy Duty Vacuum Bottle (2 pages)
- pH 700 Ion 700 (18 pages)
- pH Tutor Benchtop Meter (2 pages)
- pH 4.01 Buffer (1 page)
- pH 5+ (28 pages)
- pH 11 (76 pages)
- pH 300 (24 pages)
- 4080 TRACEABLE HUMIDITY_TEMPERATURE DEW POINT METER (2 pages)
- 4088 TRACEABLE HYGROMETER_THERMOMETER (1 page)
- 3651 TRACEABLE ANEMOMETER_HYGROMETER_THERMOMETER_DEW POINT PEN (2 pages)
- 3652 TRACEABLE ANEMOMETER_HYGROMETER_THERMOMETER_DEW POINT BAROMETER PEN (3 pages)
- Sibata 6168 Filter Holders (1 page)