Multichannel Systems STG1000 Manual User Manual
Page 80

Stimulus Generator 1000 Series User Manual
74
Operating system
Windows
© 98, ME, NT, 2000, or XP; English and German versions
supported
Data import
ASCII file format
13.2 Technical Specifications STG1004/8
Operating temperature
10 °C to 35 °C
Storage temperature
0 °C to 50 °C
Dimensions (W x D x H)
275 mm x 330 mm x 115 mm
Weight 4.8
kg
Fuse
0.2 A @ 230 V, slow blow; 0.4 A @ 115 V, slow blow
Supply voltage
Available types: 100 VAC, 115 VAC, or 230 VAC @ 50 to 60 Hz
Number of analog output
channels
4 (STG1004) or 8 (STG1008)
Output voltage
8 V to + 8V @ max. +/-20 mA
Output voltage resolution
2 mV
Ouput voltage slope
> 4 V/μs
Rise time (10-90 %)
3 to 5 μs @ delta U = 1 V or @ amplitude > 100 μA & R
L
= 10
kilohms
Output current
0.8 mA to +0.8 mA @ max. 100 V compliance voltage (between
+I and I)
Output current resolution
200 nA
Output current slope
100 μA/μs @ R
L
= 10 kilohms
Time lag between Sync Out
and voltage output
12 μs +/-2 @ amplitude > 200 mV
Time lag between Sync Out
and current output
15 μs +/-2 @ amplitude > 100 μA & R
L
= 10 kilohms)
Resolution
13 bit
Time resolution
20 μs
Output signals
Freely programmable (rectangular, ramp, sinusoidal)
Maximum frequency
(rectangular waveform)
25 kHz
Timing on external trigger
Minimum time length of
trigger input (TTL)
20 μs
Time lag between trigger
input and stimulus output
120 +/ 10 μs
Time lag between trigger
input and Sync Out
110 +/ 10 μs
Maximum trigger input
frequency
50 Hz
Interface and connectors
Sync Out for synchronization
with following devices
Digital channel via coax cable
Trigger In for synchronization
with preceding devices
Digital channel via coax cable
Interface (connection to
computer)
RS232 serial port
Download rate
56 Kbps