Mirion Technologies Genesis TLD Dosimeter User Manual
Page 2

DATASHEET
LIT4378_3_ US
TECHNICAL SPECIFICATION
Badge Name
Genesis TLD Dosimeter
Genesis Ultra TLD Dosimeter
Badge Type
16 DB= TLD 760
15 DB= TLD 760 with CR39
25 DB= TLD 760 with 115 In* and CR39
26 DB= TLD 760 with 115 In*
36= TLDMCP
35= TLDMCP with CR39
Description
4 element Harshaw TLD
(3
7
LiF:Mg, Ti [TLD700] and 1
6
LiF:Mg, Ti
[TLD600])
4 element Harshaw TLD
(3
7
LiF:Mg, Cu, P [TLD700H] and 1
6
LiF:Mg,
Cu, P [TLD600H])
Manufacturer
TLD: Thermo Electron RMP
CR39: PPG
TLD: Thermo Electron RMP
CR39: PPG
Accreditations/Approvals/Licenses
NVLAP (Code: 100555-0)
HSE (United Kingdom)
NVLAP (Code: 100555-0)
HSE (United Kingdom)
DoELAP***
Holder Type
Genesis
Genesis
Wear Location
Area, equipment, collar, lower left extrem-
ity, lower-right extremity, non-personal use,
non-specific extremity, upper-left extremity,
unknown, upper-right extremity, whole body
Area, equipment, collar, lower left extrem-
ity, lower-right extremity, non-personal use,
non-specific extremity, upper-left extremity,
unknown, upper-right extremity, whole body
Minimum Reportable Dose
10 mrem (0.10 mSv)
1 mrem (0.01 mSv)
Useful Dose Range
10 mrem - 1000 rad (0.10 mSv - 10 Gy)
1 mrem - 1000 rad (0.01 mSv - 10 Gy)
Energy Response
Beta (MAX) 0.766 MeV - 5 MeV
Photon 5 keV - 6 MeV
Neutron (TLD): Thermal - 6 MeV
Neutron (CR39): 200 keV - 6 MeV**
Beta (MAX) 0.766 MeV - 5 MeV
Photon 5 keV - 6 MeV
Neutron (TLD): Thermal - 6 MeV
Neutron (CR39): 200 keV - 6 MeV **
*Not accredited for personnel monitoring
**Neutron energies up to 20 MeV with CR39 and special calibration.
***DoELAP accreditation only applies to the Genesis Ultra with CR39
© Copyright 2008, All rights reserved. For trademark and registered trademark information.
2652 McGaw Avenue
Irvine, CA 92614
USA
Web: https://dosimetry.mirion.com
U.S./Canada +1.800.251.3331
U.K. 0170.629.9329
Worldwide +1.949.419.1000