Physical characteristics, Particulate (abpm 201), Radiation detected: alpha, beta and gamma – Mirion Technologies PING 206S User Manual

Page 2: Detector: dual large area silicon, Filter type: fslw2 (millipore), Typical measurement range: - alpha: 10, Bq/m, Μci/cc), Iodine (im 201), Radiation detected: gamma

background image

Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.

Mirion Technologies (MGPI) SA

Route d’Eyguières

FR-13113 Lamanon

France

T +33 (0) 4 90 59 59 59

F +33 (0) 4 90 59 55 18

Mirion Technologies (MGPI) Inc

5000 Highlands Parkway

Suite 150

Smyrna, GA 30082

USA

T +1 770 432 2744

F +1 770 432 9179

Mirion Technologies (MGPI H&B) GmbH

Landsberger Strasse 328a

DE-80687 Munich

Germany

T +49 (0) 89515 13-0

F +49 (0) 89515 13 169

Mirion Commercial (Beijing) Co., Ltd.

Shanghai Jiangchang Commercial Branch

Room 801, 78 Jiangchang SanLu

Zhabei District, Shanghai 200436

PR of China

T +86 21 6180 6920

F +86 21 6180 6924

PING 206S

Combination Monitors

PHYSICAL CHARACTERISTICS

Particulate (ABPM 201):

Radiation detected: alpha, beta and gamma

Detector: dual large area silicon

Filter type: FSLW2 (MILLIPORE)

Typical energy windows:
- Alpha: 2 MeV to 10 MeV
- Beta: 80 keV to 2.5 MeV
- Gamma: 80 keV to 2.5 MeV

Typical measurement range:

- Alpha: 10

-2

to 3.7 10

+6

Bq/m

3

(2.7 10

-13

to 10

-4

µCi/cc)

- Beta: 1 to 3.7 10

+6

Bq/m

3

(2.7 10

-11

to 10

-4

µCi/cc)

Iodine (IM 201):

Radiation detected: gamma

Detector: 1¼”x1” NaI(TI)

Iodine cartridge: 57.7 mm (2.27 in)

Energy range: 100 keV to 3 MeV

Typical energy window: 314 - 414 keV (

131

I, Eγ 364.5 keV)

1024 channels spectrum

Typical measurement range: 3.7 to 3.7 10

+6

Bq/m

3

(10

-10

to 10

-4

µCi/cc)

Noble gas (NGM 204):

Radiation detected: beta and gamma

Detector: dual large area silicon

Sampling chamber: 300 ml (300 cc)

Typical energy windows:
- Beta: 80 keV to 2.5 MeV
- Gamma: 80 keV to 2.5 MeV

Typical measurement range:
-

85

Kr: 3.7 10

+4

to 7.4 10

+13

Bq/m

3

(10

-6

to 2 10

+3

µCi/cc)

-

133

Xe: 3.7 10

+4

to 3.7 10

+12

Bq/m

3

(10

-6

to 10

+2

µCi/cc)

ENVIRONMENTAL CHARACTERISTICS

Long term temperature: +10°C to +40°C (+50°F to +104°F)

Periodic temperature: -5°C to +55°C (+23°F to +131°F)

MTBF: > 20 000 hours, with preventive maintenance

TID: 100 Gy (10

+4

rad)

PNEUMATIC CHARACTERISTICS

Standard fl ow rate: 35 l/min (1.24 scfm)

Pressure drop: 100 to 350 mbar (1.45 to 5.07 psi)

MECHANICAL CHARACTERISTICS

Dimensions: 1614 mm x 1535 mm x 690 mm
(63.5 in x 60.4 in x 27.1 in)

Weight: between 670 kg (1477 lb) and 720 kg (1587 lb)

Color: gray RAL 7030 (decontaminable paint)

Inlet tube connection: Ø 25.4 mm OD (1 in)

Outlet tube connection: Ø 12 mm OD (1/2 in)

ELECTRICAL CHARACTERISTICS

Power supply: refer to possible versions

Data link outputs: 1 RS232 and 5 isolated RS485

Alarm relays: 9 SPDT relays and 5 DPDT relays

I/O: 8 isolated analog outputs and 4 isolated analog
inputs (0/4-20 mA)

SIGNALING

Graphic display: measurement, historical trend, status…

Sound alarm: buzzer 90 dBA at 1 meter

Visual alarm: 3 lights (red, yellow, green)

REFERENCE STANDARDS

Nuclear: IEC60761, IEC61171, IEC61172, IEC61578

Environmental: IEC60780, IEEE323, RG 1.97

Seismic: IEC60980, IEEE344

EMC: 2006/95/CE and 2004/108/CE, EPRI 102323,
MIL STD 461 E, IEC61000-6-2 and IEC61000-6-4

VERSIONS

230 Vac or 230 Vac + 400 Vac 3Ø or 120 Vac + 400 Vac 3Ø

Solenoid check sources for ABPM 201, IM 201, NGM 204

PIS particulate and iodine samplers

Second pump for redundancy

ACCESSORIES

Local & remote display units

Calibration tools

Software

USB converters

www.mirion.com